Download document () of 20
We make what matters work*

IS300HGDV

Isolator Series Dual-Voltage 115/230V 300W 60601-1 Medical-Grade Isolation Transformer, C14 Inlet, 4 C13 Outlets

  • Protects sensitive equipment from EMI/RFI line noise and surges
  • Automatically shuts down all outlets in case of overload
  • Offers complete isolation from input AC mains power
  • Cumulative leakage current less than 50 µA
  • Ideal for use in any industrial, network or medical environment

General specifications

Product Name
Eaton Tripp Lite series isolation transformer
Catalog Number
IS300HGDV
UPC
037332199980
Product Length/Depth
12.8 in
Product Height
8.27 in
Product Width
9.84 in
Product Weight
12.35 lb
Compliances
RoHS
CE Marked
TAA Compliant
Certifications
IEC 60601-1
CSA Std. C22.2
NOM
Global Catalog
IS300HGDV
Product Type
Isolation transformer

Product specifications

Special features
Important Note Please see below under Resources > Product Literature for the Power Cord Selector Worksheet for both input and output power cord selection.
Frequency rating
50/60 Hz
Type
Medical Grade, Low Impedence Toroid- Line Isolation (Pass Through or Step UP or Step Down)
Voltage rating
115V/230V
Input Nominal Voltage Description
115/230 VAC (Selected by Input Power Module Fuse Installation) Operating Input Voltage Range 100-125Vac or 200-250VAC
Nominal Voltage / Frequency
115/230 VAC 50/60 HZ (Voltage Selected by Input Power Module Fuse Installation
Output Voltage Regulation (Line Mode)
5.2%
Operating Temperature
32° to 104°F (0° to 40°C)
Output voltage
115V/230V (Selected by Output Switch) Can also be step up or step down, Selected when Installing Fuses.
Color
White
Humidity
5% to 95% RH (Non-Condensing)
Outlets
Qty 4 Receptacles
AC Suppression Joule Rating
240
Maximum Surge Amps
70A
Input Current
3.15A @ 115V / 1.6A @ 230V
VA / Watt Capacity
300
Cooling Method
Convection
Material
Metal (Cabinet)
Noise Suppression Rating
35-65 dB
Over Current Protection
2 x 3.15A @ 115V and 2 x 1.6A at 230V ( Both Fuses included with Product as Accesories)
Leakage Current
Max AC Leakage Current 50 (uA)
Operating Elevation
0-10000 ft. (0-3000 m)
Outlet Details
Qty 4 Receptacles
Output Isolation Transformer
Toroid with AC output Floating with Respect to Earth Ground. POAG and Earth Ground Studs.
Power Input Receptacle
IEC-320-C14
Storage Temperature Range
5° to 122°F (-15° to 50°C)
System Type Selections
Input Voltage equals Output Voltage OR 120V input equals 230V output OR 230V Input equals 120V output - (Selectable when installing Fuses) Input Frequency always equals Output Frequency
THDi
Less than 1%
Total Outlets
4
TAA Compliance
Yes
Frequency Compatability
50/60 Hz
Output Receptacles
C13
Package Contents
Owner's manual
Input/output cable selector data sheet
IS300HGDV Isolator Series Dual-Voltage 115/230V 300W 60601-1 Medical-Grade Isolation Transformer
(2) Fuse sets (115V and 230V operation)
Output Receptacles Details
Compatible with Eaton Tripp Lite Power Cords

Export product specification

Need product support?

Support from Eaton

Contact the Eaton technical resource team

Phone: (800) 809-2772, option 5, option 2

Or send us an email.